Medical Equipment - Centrifuges
|
Here's a list of common problems and solutions involving Centrifuges. If you don't see the answer to your
question here, please call Precise Biomedical at 866-772-6011 and ask to speak to one of our staff engineers.
|
| Problem |
Possible Solution |
| Unit won't start |
- Safety switch on lid not actuating, check for bent switch
- Wall outlet is dead, check wall outlet with another device like a lamp or fan etc.
- Check for blown fuse on centrifuge if applicable
- Possible timer problem. If mechanical, lubricate with spray lubricant or contact cleaner
|
| Unit vibrates excessively |
- Unit is unbalanced, balance load with a counter weight opposite the tube being processed. If tube is
already in place it should have equal weight to the tube being processed; add water to counter-balance
as needed
- Possible feet missing from bottom of unit; check and replace as needed
- Unit being operated on unleveled surface. Relocate unit to level surface
|
| Unit is very noisy |
- Possible foreign body under rotor head. Empty unit and invert until foreign body is freed
- Possible bearings or brushes bad
|
|