Medical Equipment - Defibrillators
|
Here's a list of common problems and solutions involving Defibrillators. If you don't see the answer to your
question here, please call Precise Biomedical at 866-772-6011 and ask to speak to one of our staff engineers.
|
| Problem |
Possible Solution |
| Unit won't power up |
- Possible dead wall outlet--check outlet with another device such as a lamp, etc
- Possible dead battery. Charge battery for at least 12 hours and try again
- Unit charged for at least 12 hours and won't hold a charge: call Precise for replacement battery
- Possible blown fuse, check and replace fuse if applicable
- Possible bad charge button. Most machines have more than one charge button. Try the different charge button
- Possible insufficient battery power to charge capacitors. Charge batteries and try again
- Possible bad capacitors in unit. Call Precise Biomedical for additional tech support: 866-772-6011
|
| Unit won't discharge |
- Possible bad paddle switch. Place paddles in base and select discharge on selector. Call Precise
Biomedical for additional tech support: 866-772-6011
|
| Heart Monitor won't function |
- Check status of battery for heart monitor portion. Change battery if necessary. Possible new batteries
required if they don't take a change. Order new batteries from Precise BioMedical
|
| Trace is wandering or poor quality |
- Possible new PT Cable needed. Possible bad electrodes. Replace electrodes and try trace again
|
|