Medical Equipment - Sphygmomanometers
|
Here's a list of common problems and solutions involving Sphygmomanometers. If you don't see the answer to your
question here, please call Precise Biomedical at 866-772-6011 and ask to speak to one of our staff engineers.
|
| Problem |
Possible Solution |
| Unit is inaccurate |
- If unit is mercurial, remove coiled extension cord from base of unit, mercury should be at zero. If
below or above replace mercury tube
- If unit is aneroid style, make sure dial reads zero when extension tube is removed. If not replace aneroid dial
|
| Pressure bleeds off before valve is opened on bulb |
- Using a hemostat, pinch coiled extension tubing at the junction where the cuff is attached. If unit
stops leaking then, the problem is in the bladder or in the bulb. If the unit continues to leak, then
the problem is in the coiled extension tubing. Call Precise Biomedical for replacements parts: 866-772-6011
|
|