Here's a list of common problems and solutions involving Spirometer. If you don't see the answer to your
question here, please call Precise Biomedical at 866-772-6011 and ask to speak to one of our staff engineers.
|
| Problem |
Possible Solution |
| Unit won't calibrate with 3 liter syringe |
- Check for good connection between mouth piece and syringe
- Check for good connection between mouth piece and machine
- Check for cracked, stiff, or yellowed interconnect tubing. Call Precise for replacement tubing
- It is possible the 3 liter syringe is leaky or inaccurate. If possible, check unit with a different 3
liter syringe. Call Precise Biomedical for additional technical support: 866-772-6011
- If unit is battery powered it is possible a low battery condition is causing inaccurate or unreliable
readings. Charge unit for 12 hours then try 3 liter syringe test
|
|