Medical Equipment - Suction Machines
|
Here's a list of common problems and solutions involving Suction Machines. If you don't see the answer to your
question here, please call Precise Biomedical at 866-772-6011 and ask to speak to one of our staff engineers.
|
| Problem |
Possible Solution |
| Unit won't power up |
- Possible dead wall outlet. Test by plugging a different device in outlet to see if it functions
|
| No suction when unit is on |
- Possible closed regulator. Remove tubing at the regulator and open regulator for suction, if there is
still no function regulator is bad. Order a new regulator
- Possible open in system. Start at the regulator, remove tubing until suction is restored. Then, replace
the last component (such as suction canister or broken tubing)
|
|