| Unit won't calibrate with 3 liter syringe: |
- Check for good connection between mouth piece and syringe.
- Check for good connection between mouth piece and machine.
- Check for cracked, stiff, or yellowed interconnect tubing. Call Precise for replacement tubing.
- It is possible the 3 liter syringe is leaky or inaccurate. If possible, check unit with a different 3 liter syringe. Call Precise for additional technical support @ 216-351-6011.
- If unit is battery powered it is possible a low battery condition is causing inaccurate or unreliable readings. Charge unit for 12 hours then try 3 liter syringe test.
|