Here's a list of common problems and solutions involving Spirometers. If you don't see the answer to your question here, please call Precise Biomedical at 216-351-6011 and ask to speak to one of our staff engineers.
 
Problem: Possible Solution:
Unit won't calibrate with 3 liter syringe:
  • Check for good connection between mouth piece and syringe.
  • Check for good connection between mouth piece and machine.
  • Check for cracked, stiff, or yellowed interconnect tubing. Call Precise for replacement tubing.
  • It is possible the 3 liter syringe is leaky or inaccurate. If possible, check unit with a different 3 liter syringe. Call Precise for additional technical support @ 216-351-6011.
  • If unit is battery powered it is possible a low battery condition is causing inaccurate or unreliable readings. Charge unit for 12 hours then try 3 liter syringe test.

Precise Biomedical Inc.  |  2915 Brookpark Rd.  |  Parma OH 44134
contact@precisebiomedical.com  |  216-351-6011, fax 216-351-8334